Aesculap Power Systems
Acculan® 3Ti charger GA677
4.
Working with the
Acculan® 3Ti charger

4.1 System set-up

Connecting the power supply

Prior to connecting the product to mains power:
– Check the mains power cord for any damage.
– Check the product for any damage (e. g. bent,
short-circuited contacts of the charging bays).
Plug in the mains power cord at device socket 5.
Plug in the mains plug at the mains power socket.

4.2 Function checks

Carry out a visual inspection.
Prior to connecting the product to mains power:
– Check the mains power cord for any damage.
– Check the product for any damage (e. g. bent,
short-circuited contacts of the charging bays).
Plug in the mains power cord at device socket 5.
Plug in the mains plug at the mains power socket.
6
As soon as the product is connected to mains power, it
will perform a self-test for approx. 3 seconds. During
this period all indicator elements at the front panel of
the charger must be illuminated. With the self-test
completed, only "Power On" indicator 3 is illuminated.
The product is ready for operation.
The charging process is started by correctly inserting a
fault-free rechargeable battery in a charging bay. The
respective charging progress indicator is illuminated.
Note
For any deviations from the sequence described above,
see chapter Troubleshooting list!