Battery Temperature; Charging Time; Preparation And Setup - B. Braun Acculan 3Ti GA677 Instructions For Use Manual

Aesculap power systems
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1

Battery temperature

Each charging bay is equipped with a fan, which works
independently of the battery temperature.
The charging of the battery is aborted as soon as a bat-
tery temperature of > 47 °C is detected.
Excessive battery temperature is indicated by the
"Action request" symbol illuminated in indicator field 2.
When the battery temperature drops below 45 °C, the
"Action request" symbol in indicator field 2 still
remains illuminated. The battery must be removed
from the respective charging bay and plugged in again
to restart charging.

Charging time

Charging is stopped as soon as the maximum charging
time is reached.
The permanent monitoring of the battery also allows
detecting faults in the battery block.
When a battery fault or a charging fault is detected,
this is indicated by the "Action request" symbol illumi-
nated in indicator field 2 and charging is aborted.
The charging bays can be ordered separately, see chap-
ter Accessories/Spare parts.
3.

Preparation and setup

Non-compliance with these rules will result in com-
plete exclusion of liability on the part of Aesculap.
Observe the following regulations when setting up
and operating the product:
– national regulations for installation and opera-
tion,
– national regulations on fire and explosion pro-
tection.
– operating advisories according to IEC-/VDE reg-
ulations.
Fire and explosion hazard!
Do not operate the product in explo-
sion-hazard areas.
DANGER
Malfunction of, or damage to the prod-
uct caused by improper setup!
Do not expose the product to direct
CAUTION
sunlight or moisture.
Do not cover ventilation slots 7 of
the product.
Only connect the product to a mains
power system with an equipment
grounding conductor and with a volt-
age rating 100 V to 240 V and a
mains power frequency range of
50 Hz to 60 Hz.
Ensure that the mains power socket
used for supplying the product is
freely accessible.
5

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents