Acculan® 3Ti charger GA677

5.1 Cleaning/Disinfecting

Note
Always follow disposal and hygiene instructions and
regulations!
Manual cleaning, disinfecting
Risk of electric shock and fire hazard!
Unplug the unit before cleaning.
Make certain that no fluids will pen-
DANGER
etrate the product.
Leave the product exposed to air for
at least 1 minute after cleaning/dis-
infecting.
Damage to the product due to incorrect
cleaning or inappropriate cleaning/disin-
fecting agents!
CAUTION
Only
use
agent approved for surface cleaning
according to manufacturer's instruc-
tions.
Never immerse the product in water
or detergent.
Do not introduce any fluid into the
charging bays.
Clean the contacts in the charging
bays, applying extreme care.
Carry out a wipe disinfection by wiping the product
housing with a lint-free cloth moistened with a
disinfecting agent (e. g. B. Braun Meliseptol).
If necessary, after the appropriate exposure time,
wipe off any residues of the cleaning/disinfecting
agent with a lint-free cloth moistened with clean
water.
Dry the product with a clean, lint-free cloth.
Clean the contacts in the charging bay with isopro-
panol or ethyl alcohol and a cotton swab. Do not
use corrosive chemicals.
10
a
cleaning/disinfecting
If necessary, repeat the cleaning/disinfection pro-
cess.
Control, care and inspection
Inspect the product after each cleaning and disin-
fecting cycle to be sure it is clean, functional and
undamaged.
6.

Servicing/Inspection

Send in the Acculan® 3Ti charger with its accesso-
ries for annual inspection by the manufacturer.
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone:
+49 7461 95-2700
Fax:
+49 7461 16-2887
E-Mail:
ats@aesculap.de
Or in the US
Aesculap Inc.
Attn. Aesculap Technical Services
615 Lambert Pointe Drive
Hazelwood, MO 63042
Aesculap Repair Hotline
Phone:
+1 800 214-3392
Fax:
+1 314 895-4420
Other service addresses can be obtained from the
address indicated above.