Symbols; Troubleshooting - MABIS 18-607-000 Instruction Manual

Deluxe instant ear thermometer
Table of Contents

Advertisement

Available languages

Available languages

Symbols

Fahrenheit/Celsius
Memory symbol and
measurement number
Scanning in progress
Low battery
Happy face
Indicates temperature is lower
than 100.4°F (38.0°C)
Sad face
Indicates temperature is
100.4°F (38.0°C) or higher
10

Troubleshooting

Display Symbol
Condition/Cause
Measurement is below
normal temperature range.
Measurement is above
normal temperature range.
Operating temperature is
outside of the permissible
range (60.8°F – 95.0°F;
16.0°C – 35.0°C).
Blank display
Low battery
Low battery warning
Corrective Action
Retake temperature
positioning the probe
snuggly into the ear
canal.
Retake temperature
positioning the probe
snuggly into the ear
canal.
Take thermometer to a
room where the
temperature is within the
permissible range.
Wait 45 minutes to
stabilize thermometer.
Retake temperature
positioning the probe
snuggly into the ear
canal.
Check battery polarity.
May need to replace
battery.
Replace battery
immediately.
Unit will work; however,
replace battery when
possible.
11