Run A Check Standard - Hach PHC28101 User Manual

Ph refillable probe for low ionic strength samples
Table of Contents

Advertisement

1. Connect the
2. For the best
probe to the
stabilization time,
meter. Make sure
condition the
that the cable
probe in the
locking nut is
sample for at least
securely
10 minutes before
connected to the
the initial sample
meter. Turn the
measurement.
meter on.
6. Repeat steps
7. When
2-4 for additional
measurements
measurements.
are done, store
the probe (refer to
Storage
on page 10).

Run a check standard

Measurement procedure:
3. Rinse the
probe with
deionized water
and then with the
sample. Blot dry
with a lint-free
cloth.
The run check standard feature validates instrument performance between sample
measurements. Use the run check standard feature for periodic or user-defined interval
measurements of a traceable standard solution. Set the criteria for check standards from
the PHC281 Settings menu.
Note: Access control must be off or a valid password must be entered before any of the check
standard method options can be changed.
1. Push
. The Full Access Options menu is shown.
2. Select Run Check Standard.
Note: Select the correct probe if two probes are connected to the meter.
3. Get the standard solution shown on the display.
4. Rinse the probe with deionized water. Blot dry with a lint-free cloth.
5. Put the probe in the standard solution until the temperature sensor is completely
submerged. Move the probe up or down or gently tap on the beaker to remove air
bubbles from the probe.
4. Put the probe
5. Push Read.
in the sample and
The display will
stir gently. Make
show "Stabilizing"
sure that the
and a progress
reference
bar as the probe
junctions are
stabilizes in the
completely
sample. The
submerged. Do
display will show
not put the probe
the lock icon when
on the bottom or
the reading
sides of the
stabilizes.
container. Shake
the probe from
side to side in the
sample to refresh
the reference
junction.