Calibration Procedure - Hach PHC28101 User Manual

Ph refillable probe for low ionic strength samples
Table of Contents

Advertisement

Calibration
Before calibration:
The probe must have the correct service-life time stamp. Set the date and time in the meter before the probe is attached.
It is not necessary to recalibrate when moving a calibrated probe from one HQd meter to another if the additional meter is
configured to use the same calibration options.
To view the current calibration, push
If any two probes are connected, push the UP or DOWN arrow to change to the single display mode in order to show the
Calibrate option.
Prepare the probe for use (refer to
1. Connect the
2. Push
probe to the
Calibrate. The
meter. Make sure
display shows the
that the cable
buffers that are
locking nut is
necessary for
securely
calibration.
connected to the
meter. Turn on the
meter.
, select View Probe Data, then select View Current Calibration.
Preparation for use
Calibration notes:
pH buffers can be used in any order. Use buffers that are two pH units apart.
Additional standard sets along with the minimum number of calibration points can be
selected in the Calibration Options. For a two point calibration, it is recommended that
two buffers be selected: one with a pH above and one with a pH below the expected
sample pH. For a one point calibration, select the buffer nearest to the expected
sample pH.
The calibration is recorded in the probe and the data log. The calibration is also sent
to a PC, printer or flash memory stick if connected.
Air bubbles under the sensor tip when submerged can cause slow response or error
in measurement. If bubbles are present, gently shake the probe until bubbles are
removed.
If a calibration error occurs, refer to

Calibration procedure:

3. Prepare the
fresh buffers in
separate beakers
or appropriate
containers.
on page 2).
Troubleshooting
4. Rinse the
probe with
deionized water.
Blot dry with a lint-
free cloth.
on page 11.
5. Put the probe
in the pH buffer
solution and stir
gently. Make sure
that the reference
junctions are
completely
submerged.
Shake the probe
from side to side
in the standard
solution to refresh
the reference
junction.