SPECIFICATIONS

S
P
STANDARD SPECIFICATIONS
E
Electrical Requirements:
C
Rated Current Input:
Sound Level:
I
F
I
Operating Altitude:
C
A
T
Ambient Temperature Range:
I
*Oxygen Output Concentration
O
Levels:
N
S
Maximum Outlet Pressure:
Sens O2 Alarm Thresholds:
If no internal selector switch, 73%
operation ONLY.
Flow Range:
Power Consumption:
Pressure Relief Mechanism
Operational at:
SPECIFICATIONS
120 VAC +10% (132 VAC/108 VAC), 60 Hz
4.0 A (ALL Models)
52 dB Average (3LX)
53 dB Average (5LX, 5LXO2, 5LXO2T, 6LXO2, 6LXO2T)
Up to 6,000 ft. (1828 meters) above sea level without degradation of
performance. From 6,000 ft. (1828 meters) to 13,129 ft. (4000 meters) below
90% efficiency.
50
O
F - 95
O
F (10
O
- 35
O
C)
ALL MODELS
95.6% to 87% at all flowrates.
* Stated performance maximum achieved after 30 min. of run time.
(5LX, 5LXO2, 5LXO2T, 6LXO2, 6LXO2T)
5 psi + 0.5 psi.
(34.5 kPa + 3.45 kPa.)
O2 Purity
Over 85% (+ 2%)
73% (+ 3%) to 85% (+2%)
Below 73% (+ 3%)
1/2 - 5 L/min. (5LX, 5LXO2, 5LXO2T, 6LXO2, 6LXO2T) 1/2 - 3 L/min. 3LX)
For flow rates less than 1/2 L/min., use the Invacare Pediatric Flowmeter
PF16.
460W (6LXO2, 6LXO2T) 350W/400W (3LX)
400W (5LX, 5LXO2, 5LXO2T)
35 psi + 3.5 psi (241 kPa + 24.1 kPa)
8
(3LX)
5 psi + 0.5 psi.
(34.5 kPa + 3.45 kPa.)
Indicators
Internal Switch
*Set at 73%
GREEN Indicator Light
YELLOW Indicator Light
RED Indicator Light -
Continuous Audible
TM
Alarm Sieve GARD
Compressor Shutdown
* Factory Preset at 73%
Internal Switch
Set at 85%
GREEN Indicator Light
RED Indicator Light -
Continuous Audible
TM
Alarm Sieve GARD
Compressor Shutdown
N/A