PROCEDURE 12
PROCEDURE 12 - REPLACING TRANSFORMER ASSEMBLY (FIGURE 1)
T
R
TO PREVENT ELECTRICAL SHOCK, ALWAYS DISCONNECT CONCENTRATOR FROM
A
ELECTRICAL OUTLET BEFORE SERVICING.
N
1. Unplug unit.
S
F
2. Remove cabinet (Refer to PROCEDURE 1 - REMOVING CABINET).
O
3. Remove control panel (Refer to PROCEDURE 9 - REMOVING CONTROL PANEL).
R
4. Locate the transformer assembly in the upper right corner on the back of control panel.
M
5. Unplug the transformer connector from the Control (bottom) P.C. Board.
E
R
6. Remove the two (2) keps nuts that secure transformer to the back of the control panel.
7. Reverse STEPS 5-6 to install NEW transformer.
8. Torque keps nuts to 13.5 + 1 in. lbs.
9. Reinstall control panel (Refer to PROCEDURE 9 - REMOVING CONTROL PANEL).
10. Reinstall cabinet (Refer to PROCEDURE 1 - REMOVING CABINET).
Transformer Assembly
Keps Nut
Transformer
Connector
Keps Nut
FIGURE 1 - REPLACING TRANSFORMER ASSEMBLY
CAUTION
Transformer P.C. Board Connector
(All Models)
50
TRANSFORMER
Control Panel
Control (Bottom) P.C. Board
(Typical)