34
Mechanical cleaning and disinfection of the
external coolant tubes
Use the W&H adaptor kit REF 07233500
for the thermo washer disinfector.
Screw the W&H adaptor into the adaptor on the
injector rail.
Screw the W&H intermediate adaptor onto the
W&H adaptor.
Put the W&H silicone hose over the W&H
intermediate adaptor.
Insert the external coolant tube into the W&H
silicone hose.