W&H S-11 L G Instructions For Use Manual page 33

Surgical handpieces with mini led plus and generator/without light
Hide thumbs Also See for S-11 L G:
Table of Contents

Advertisement

Mechanical cleaning and lubrication
W&H recommends mechanical cleaning and lubrication with W&H Assistina.
> Follow the instructions in the Assistina Instructions for use.
S-10, S-12: only with W&H Assistina 301
Mechanical cleaning and disinfection internal and external
The handpiece can be cleaned and disinfected in a thermo washer disinfector.
> Use the W&H adaptor kit for the thermo washer disinfector to prepare the external coolant tubes.
> Note the manufacturer's specifications on equipment, cleaning agents and detergents.
> Make sure, that the handpiece is completely dry internally and externally after thermo washer
disinfection. Remove any liquid residues with compressed air.
> Lubricate the dry handpiece immediately after thermo washer disinfection.
33