MIN limit
MAX limit
MIN limit
Alarm value
Filter time
Calibration: 4 mA
Calibration: 20 mA
9. Maintenance
Warning!
Electric shock hazard when housing is open!
•
Switch off the power supply during all maintenance and repair work on the measuring system.
•
The system may not be in operation for a sensor exchange.
Repair and maintenance work may only be carried out by qualified electricians
•
The system is maintenance-free.
•
Minimum current output value to be set
(Standard 3 mA) Range 0 ... 22 mA
Maximum current output value to be set
(Standard 20 mA) Range 0 ... 22 mA
Minimum current output value to be set
(Standard 3 mA) Range 0 ... 22 mA
Output value to be set for alarm (sensor or internal alarm),
at the same time the relay drop.
(Standard 3 mA) Range 0 ... 22 mA
Filter time to be set for the current output.
(Standard 1 s) Range 0.1 ... 999.9 s
Setting of the min. starting current. Adjustment to the external
measuring system (in case of deviating display). Adjust the
current output to 4 mA with the < and > buttons.
Setting of the max. starting current. Adjustment to the external
measuring system (in case of deviating display). Adjust the
current output to 20 mA with the < and > buttons.