8. Setting Parameters

To set the time parameters (DURA, DELAY or INTER), first press the selected
parameter key and then the desired time.
Example: To set the duration of channel 6 to be 52 msec (52 x 10
is given in seconds).
Press keys: 'DURA, 6' - the previous duration is displayed.
(If the duration time has never been set, the sign 'FFFF FF' is displayed).
Press keys: '52, ENTER, 3, ENTER' ('3' for msec).
52 msec can be inserted also as 0.052 x 10
or 5200 x 10
For your convenience it is recommended to use only exponents of 6 (µsec), 3 (msec) or
0 (sec).
Thus, for 52 msec. use either 52 msec. or 0.052 sec.
Note:
The new parameter becomes operative the instant you press the last ENTER.
8.1
Using the '
You can increase or decrease each time parameter using the '↑' and '↓' keys.
Example:
Press keys: 'DURA, 6'.
Hold down the '↑'or '↓' keys to change the duration of channel 6.
Note: The change takes place only when releasing the key.
Press key: 'CLEAR-DISPLAY' in order to clear the display.
8.2

Setting the 'M' Parameter

Example: To set M=80 pulses per train for channel 8 -
Press keys: 'M, 8' - The previous M8 is displayed.
Press keys: '80, ENTER, 0, ENTER'.
Note:
Only integer numbers are accepted for M.
The exponent should equal 1 only for M > 9999.
-5