Wireless Compliance Information For Eu Directive 1999/5/Iec; Grid Alignment Intended Use - Carestream DRX-Revolution Safety And Regulatory Information Manual

Mobile x-ray system
Hide thumbs Also See for DRX-Revolution:
Table of Contents

Advertisement

Safety and Regulatory Information
Wireless Compliance
Information for EU
Directive 1999/5/IEC
Grid Alignment Intended
Use
1-26
• DLink DAP 2553 Access Point
• RF Ideas Model RDR–80_82AKU Proximity Badge Reader
• Bar Code Reader (Code Corp. Radio Modem,
Model CR2AG–R0–01–02)
• Kensington BlueTooth USB Micro Adapter, Model M01011
• Intel Centrino Ultimate N 6300 WLAN module, Model 633AN
Refer to the appropriate manufacturer documentation for wireless regulatory
declarations.
The CARESTREAM DRX-Revolution Mobile X-ray System uses the same
vendor, type, and model antennae originally certified with the Intel 6300
WLAN module and the DLink DAP 2553 Access Point.
When the user obtains an X-ray image, there is generally an optimal alignment
between the radiation source and the two-dimensional receptor that records
the image data. In most cases, the X-ray source provides radiation in a
direction that is perpendicular to the receptor. Use of an anti-scatter grid can
significantly improve the quality of X-ray images. When using a grid, the
requirements for good alignment are much more stringent. For mobile X-ray
systems, the receptor is placed behind the patient and is blocked from the
operator's view by the patient, bedding, and so forth. This makes it very
difficult to align the X-ray source to the detector properly. The grid helps an
operator achieve the correct alignment.
Principle of Operation
The Grid Alignment System uses transmitter emitter coils mounted on the
X-ray source to produce a rotating magnetic field vector.
Magnetic field sense coils are mounted in the perimeter of a frame that holds
the grid and DRX-1 detector. The signal from the emitter is seen by the sense
coils. The processor on the cart interprets the signals to determine the
relative alignment of the X-ray tube and the detector, and displays it on a
graphical user interface.