Electromagnetic
Emissions/Immunity
Port/Cable
USB
Ethernet
Ethernet
Prep/Expose Hand Switch
AC Power Cord
Electromagnetic Compatibility Precautions
1.
Medical electrical equipment requires special precautions regarding
electromagnetic compatibility (EMC). Medical equipment must be
installed and put into service according to the EMC information
provided.
Communications Equipment
2.
Portable and mobile radio frequency (RF) communications equipment
can affect medical electrical equipment EMC performance.
Replacement of cables, or accessories, other than those specified below
3.
with the exception of transducers or cables sold by the equipment
manufacturer as replacement parts for internal components, may result
in increased emissions or decreased immunity of the medical equipment.
Cable and accessory information for the System:
4.
Cable Information for the System
1
BlueTooth
2
Keyboard/Mouse/DVD
3
Keyboard/Mouse/DVD
1
Hospital Uplink
1
Tethered Detector Link
1
Prep/Expose/Collimator Light
Controller
1
Retractable
Shielded Locations
5.
The System is fully compliant with the requirements of
IEC 60601–1–2: 2007 without being used in a shielded room.
9J0239
Safety and Regulatory Information
Function
Customer
Service
Service
Customer
Customer/
Service
Customer
User
Cable
Length
NA
1 Meter
1 Meter
3 Meter
3 Meter
3 Meter
3 Meter
1-19