Safety and Regulatory Information
Electromagnetic Immunity for Non-Life Supporting Equipment and Systems
a
Field strengths from fixed transmitters, such as base station for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the System is used exceeds the
applicable RF compliance level above, the System should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating the System.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 v/m.
9J0239
1-23