Current Designs Button Box and Trigger Setup (FORP 932)

There is no on/off switch and this device does not need to be unplugged

Device Selection

1. DO NOT set the interface settings until you plug in the button box inside the scanner
room – the button responses will not work
2. Note - you do not need to plug in a box if you only need a trigger – just choose your
interface setting with appropriate trigger (5s or Ts)
3. Click dial in once and scroll to yes in "change modes" menu and click dial once
4. Scroll to "autoconfigure" and click dial once
5. Scroll to "usb" and click dial once
6. Below is a list of available devices – autoconfigure will automatically choose the box that
is connected in the scanner room
HHSC - 1X4-L
HHSC - 2X2
o Left box sends dnwe or 9876
o Right box sends bygr or 1234
HHSC – 2x4
 HHSC Joy-1 joystick (older model than what is on the current designs website)
HHSC TRK-1 trackball
7. You may see other devices listed that are not available, contact BMC Techs for more info
8. Once you have selected your device, you will see the "Response Settings" menu

Response Settings

1. Scroll to applicable setting and click dial once
 HID KEY BYGRT - use this for letter responses and T is the trigger
o this is the same as 0 setting on previous interface box
 HID KEY 12345 - use this for number responses and 5 is the trigger
o this is the same as 4 setting on previous interface box
 HID KEY NAR - Non-auto release – (release code is not sent until key is
released)
o this is the same as 2 setting on previous interface box
o
 HID KEY NAR 12345 - Same as above but with numbers
 HID KEY 1-9 No 5
o For MacStim users – no trigger will be sent - users must manually
28
- buttons straight in a row
- split boxes, you may use just one box if applicable
- curved row of buttons, you may use just one box if applicable
this setting is often required for psycho tool box in matlab
start scan
Revision Date: 3/13/2019