Fluke i400 Calibration Information Manual page 6

Ac current clamp
Hide thumbs Also See for i400:
Table of Contents

Advertisement

i400
Calibration Information
Current Accuracy Tests
1. Connect the equipment as shown in Figure 2.
2. Center the UUT jaws around the single-turn #14 gauge wire.
3. Set the 5520A output as called out in Table 3. Check that the Multimeter readings are within the
specified limits in step 1 of Table 3.
4. Place the 5520A in Standby mode.
5. Connect the equipment as shown in Figure 3.
6. Center the jaws of the UUT around the 5500A coil.
7. Apply the 5520A output as called out in Table 2, steps 2 and 3. Verify that the Multimeter readings are
within the specified limits.
Step
Nominal Value
1.
1 A, 45 Hz
2.
100 A, 400 Hz
3.
400 A, 60 Hz
*To obtain the best TUR, the Fluke 189 is pre-calibrated at each step using the 5520A with LCOMP OFF. The Fluke 189
error is then algebraically added to the UUT reading. The Fluke 189 must be in a mA B function.
6
Table 3. Accuracy Tests
5520A LCOMP
5520A Output
Setting
1 A, 45 Hz
OFF
2.0 A, 400 Hz
ON
8.0 A, 60 Hz
ON
Figure 2. Connections for Current-Accuracy Test (Step 1)
*Multimeter Reading limits
UUT Nominal
Output
Lower
1 mA, 45 Hz
0.920 mA
100 mA, 400 Hz
97.94 mA
400 mA, 60 Hz
391.94 mA
Upper
1.080 mA
102.06 mA
408.06 mA
189
LOGGING MULTIMETER
AutoHOLD
FAST MN MX
LOGGING
YES
HOLD
HOLD
MIN MAX
MIN MAX
REL
REL
CANCEL
SAVE
NO
Hz
Hz
%
%
ms
ms
RANGE
RANGE
SETUP
nS
nS
F
F
mV
mV
C
C
ac+dc
ac+dc
A
A
V
V
mA
mA
ac+dc
ac+dc
mV
mV
A
A
dB
dB
A
A
V
V
dB
dB
mA
mA
ac+dc
ac+dc
A
A
OFF
OFF
ac+dc
ac+dc
VIEW
CLEAR MEM
TEMPERATURE
TEMPERATURE
mA
mA
A
A
COM
COM
V
V
A
A
bbs04f.eps

Advertisement

Table of Contents
loading

Table of Contents