Replacing the Fuses
replacement fuses ∅ ∅ 5x20 0.5A/250V FAST or ∅ ∅ 6x25
To avoid electrical shock or damage to the meter, only use
10A/250V FAST.
Referring to Figure 19, examine or replace the meter's fuses as
follows:
1. Turn the Function Rotary Switch to the OFF position and remove
the test leads from the terminals.
2. Remove the holster from the meter. Then, using a screwdriver,
unscrew the two screws on the back case and remove it.
3. Remove either fuse by gently prying one end loose, then sliding
the fuse out of its bracket.
voltage, and speed ratings: ∅ ∅ 5x20 0.5A/250V FAST or ∅ ∅ 6x25
4. Install ONLY the specified replacement fuses with the amperage,
10A/250V FAST.
5. Reinstall the back case. Secure the back case by turning the
screws clockwise.
Screw
∅6x25 10A/250V FAST
Elenco
®
part # 531101
∅5x20 0.5A/250V FAST
Elenco
®
part # 533004
All manuals and user guides at all-guides.com
Warning
!
Figure 19. Fuse Replacement