Agilent Technologies AssayMAP Bravo Installation Manual page 104

Platform
Table of Contents

Advertisement

Verifying the installation and setup
Performing a mock run
Protocol step
Internal Cartridge
Wash 1
Cup Wash 2
Internal Cartridge
Wash 2
Stringent Syringe
Wash
Elute
Final Syringe Wash
96
Head moves to
Bravo Platform action
deck location...
5
Aspirates Cartridge Wash Buffer 1 into the syringes.
2
Mounts the cartridges on the head.
1
Dispenses 5 µL Cartridge Wash Buffer 1 through the
cartridges at the Load Samples flow rate for the Sample Chase
step.
1
Dispense Cartridge Wash Buffer 1 through the cartridges for
the Internal Cartridge Wash step.
1
Wash the exterior of the cartridge tips at the wash station.
2
Park the cartridges at the seating station.
1
Wash the syringes at the wash station.
6
Aspirate the Cartridge Wash Buffer 2 into the syringes.
2
Performs Cup Wash.
1
Dispenses the buffer at the wash station.
1
Washes the syringes at the wash station.
6
Aspirates the Cartridge Wash Buffer 2 into the syringes.
2
Mounts the cartridges on the head.
1
Dispenses the Cartridge Wash Buffer 2 through the cartridges
for the Internal Cartridge Wash 2 step.
1
Washes the exterior of the cartridge tips at the wash station.
2
Parks the cartridges at the seating station.
1
Washes the syringes at the wash station.
8
Aspirates the Stringent Syringe Wash Buffer (Elution Buffer).
1
Dispenses the buffer at the wash station.
1
Washes the syringes at the wash station.
8
Aspirates the Elution Buffer.
2
Mounts the cartridges.
9
Elutes the samples into the Eluate Collection plate.
1
Washes the cartridge exteriors at the wash station.
2
Parks the cartridges at the seating station.
9
Mixes the eluates.
1
Washes the syringes at the wash station.
1
Washes the syringes at the wash station.
AssayMAP Bravo Platform Installation Guide

Advertisement

Table of Contents
loading

Table of Contents