Directions for use
Protection against electrosurgical
Electrosurgical protection per EN60601-2-27:2011. ECG device will resume as per the
table below.
Parameter
HR value
ECG waveform
Respiration value
Pacer detect
Arrhythmia
Device classification
ECG device will operate per table below after startup and after changes to parameter
configurations (e.g., filter on/off) or lead off recovery.
Parameter
HR value
ECG waveform
Respiration value
Pacer detect
Arrhythmia
Default settings
Characteristic
Lead color code
Lead for view
Sweep speed
Initial report after
Maximum time to
defibrillation
recover
Acquiring
10 sec +5 beats
Waveform
≤10 sec
Acquiring
10 sec +5 breaths
Pacer marker in waveform
≤10 sec
Acquiring or
10 sec +5 beats
Cannot analyze
Initial report after
Maximum time to
defibrillation
recover
Acquiring
10 sec +5 beats
Waveform
≤3 sec
Acquiring
10 sec +5 breaths
Pacer marker in waveform
≤3 sec
Acquiring or
10 sec +5 beats
Cannot analyze
Specification
AHA
II
25 mm/second
Specifications 33
Note
Depends on patient heart
rate
Depends on patient
respiration rate
Depends on patient heart
rate
Note
Depends on patient heart
rate
Depends on patient
respiration rate
Depends on patient heart
rate