14 About warnings and cautions
Welch Allyn Connex
Devices ECG Module
WARNING Personal/patient injury risk. Wall Mounts must be installed by
authorized Welch Allyn service personnel or a Biomedical engineer to
ensure mounting integrity and proper placement.
WARNING Personal/patient injury risk. Only authorized Welch Allyn
service personnel or a Biomedical engineer should attach or remove the
device from the wall mount.
WARNING Personal/patient injury risk. Any modification made to a Welch
Allyn mounting solution removes Welch Allyn from responsibility or liability
and voids the warranty.
WARNING Personal/patient injury risk. Welch Allyn is not responsible for
the integrity of any installation not performed by authorized Welch Allyn
service personnel.
WARNING Personal/patient injury risk. Welch Allyn is not responsible for
the integrity of any wall structure or wall mounting interface. Welch Allyn
recommends that you contact your Biomedical Engineering Department or
maintenance service to ensure professional installation, safety, and
reliability of any mounting accessory.
CAUTION Position the wall mount so that the screen, controls, and
connectors are accessible and support optimal and ergonomic use of the
device.
CAUTION Never use acetone, ether, freon, petroleum derivatives, or other
solvents to clean the ECG Module. Never immerse the ECG module or the
patient cable in liquid. Never autoclave or steam clean the ECG module or
the patient cable. Never pour alcohol directly on the ECG module or the
patient cable, and never soak any components in alcohol. If any liquid
enters the ECG module, remove the ECG module from service, and have it
inspected by a qualified service person before using it again.
CAUTION Verify that dates on applicable accessories have not expired.
CAUTION Connect the USB cable to the Connex device in a manner that
minimizes entangling.
CAUTION To prevent accidental loosening of the USB cable and the
potential loss of ECG connection to the device, you must close the door
and tighten the screw to secure the cable in place.