Welch Allyn Connex ECG Module Directions For Use Manual page 35

Table of Contents

Advertisement

Directions for use
Characteristic
Mains filter
Frequency range
Power source
Digital data interface
Monitoring period
ECG channels
Sweep speed
Resolution
Lead-off detection
Applied currents
User-adjustable parameters
ECG cable specifications
Characteristic
Connector at ECG module
ECG cables
Patient cables
Impedance respiration specifications
Characteristic
Impedance Respiration accuracy
Range
Breath detection range
Respiration lead source
Respiration, leads-off sensing and active
noise suppression (applied currents)
Specification
50 Hz, 60 Hz, off (default = 60 Hz)
0.5 to 70 Hz
USB (4.5 V to 5.5 V)
Serial (USB – Full Speed)
Continuous
Supports 3-lead or 5-lead
25 mm/sec, 50 mm/sec
14 bit ECG data, 30 mV peak-to-peak dynamic range minimum
3-lead or 5-lead
Lead connection failures are detected by a bias current with respect to
the reference electrode.
See Alarms
Specification
80 in. ±1 shielded cable to a 24 in. ±1 lead fan-out to electrode
connection
3-lead or 5-lead to snap connection
Conforms to ANSI/AAMI EC53
Specification
±2 breaths per minute or ±2%, whichever is greater
5 to 100 breaths per minute
0.4 ohms to 3.0 ohms
Lead II (right arm and left leg)
Note
Respiration can only be acquired from Lead II.
Respiration – Excitation signal characteristics Less than 25 uA RMS at
31 kHz pseudo sine wave
Lead off – DC current of 50nA max for RA ,LA, LL, V; 200nA max for RL
Specifications 31