Mount And Connect The Ecg Module - Welch Allyn Connex ECG Module Directions For Use Manual

Table of Contents

Advertisement

16 Setup
No. Feature
3
LED module status indicator
4
ECG patient cable connector
5
ECG patient cable
6
Mounting clip

Mount and connect the ECG module

To mount the ECG module on the desired mounting apparatus (basket, bin, wall
mount), follow the instructions presented in the "Connex® Devices ECG Module
Assembly Instructions" packaged with the module.
Note
Note
CAUTION Mount the ECG module before use. Position the
module as instructed using only the provided mounting
fixture. Any other mounting position or setup will
compromise water ingress protection.
CAUTION Connect the USB cable to the Connex device in a
manner that minimizes entangling.
CAUTION To prevent accidental loosening of the USB cable
and the potential loss of ECG connection to the device, you
must close the door and tighten the screw to secure the
cable in place.
CAUTION Accessories attached to this device must run on
battery power. Do not use any accessory's external power
supply when it is attached to the monitor.
On some stands, the mounting bracket partially overlaps the
USB door. If you observe this overlap, loosen the screw on
the mounting bracket and shift the device forward just
enough on the bracket to open the USB door, then slide the
device back on the mounting bracket.
If you loosened the mounting bracket in order to open the
USB door, shift the device forward on the bracket, close the
door, and then slide the device back into its mounted
position. Ensure that the device is seated firmly on the
bracket, then tighten the screw on the bracket into the
screw hole on the back of the device. (See "Mount the
monitor on a stand" in the Welch Allyn Connex® Devices
directions for use for more detail.)
Welch Allyn Connex
Description
Indicates to module status as follows:
No LED: Off, not powered
Steady green: Powered
Flashing green: Powered, enumerated, sending patient data
Steady yellow: Powered, internal error or POST
Provides connection for an ECG patient cable with either 3 or 5
leads to snap connectors
Connects to the module and to either 3 or 5 leads to snap
connectors
Provides the surface onto which the back of the module slides to
mount the module on a bin or basket
®
Devices ECG Module