1. Safety Precautions 1.1 Using this manual This manual is for the users who operate the device for the first time. This manual provides information on the detailed instructions, precaution, troubleshooting and maintenance care. Keep this manual nearby for future reference. 1.2 Safety Labels on the Device The labels on the device represent safety instructions and directions.
6. Should not use flammable, toxic, radioactive, explosive or corrosive materials. 7. The containers should be kept level while drying the samples in them. 8. Any kind of device repair must be performed only by Hanil Scientific Inc. authorized service engineers.
2. Product Description 2.1 Product Description [Back side of the device] ① ② ③ ④ [Right side of the device] ⑤ ⑥ ⑦ ① Vacuum Release Valve ② Tube Connection to Vacuum Pump ③ Mains Inlet ④ Power Outlet for Vacuum Pump ⑤...
2. Product Description 2.2 Delivery Package · Check all parts for damage that may have occurred during transportation. · Check the delivery for completeness. - Hyper COOL 3055/3110 - AC Power Cable - Hose - Operation Manual - Clamp 7 / 26 Technical Support : +82-2-3452-8966 / techsupport@ihanil.com...
3. Installation 3.1 Wheel Lock 1. Press the lever of the swivel caster to lock the wheel. 3.2 Switching On the Device · After installing the device, lock the wheel to prevent the device from moving. 1. Connect the AC power cord to the power socket on the backside of the device and put the plug into a power outlet.
3. Installation 3.3 HyperCOOL Connection [HyperCOOL as a 'Freezer Dryer'] ① ⑦ ⑤ ② ⑥ ④ ⑤ ③ ① Rubber Valve Closed ② Tube Connection to Vacuum Pump ③ Mains Inlet ④ Vacuum Pump to Power Outlet ⑤ Vacuum Pump to HyperCOOL ⑥...
Page 11
3. Installation 3.3 HyperCOOL Connection [HyperCOOL as a 'Freezer Dryer'] [Preparation of Chamber] 1. Ensure all the vacuum valves are firmly closed. 2. Ensure all the connections and attachment from the chamber are tightly sealed. 3. Ensure the drain valve does not contain any water. 4.
Page 12
3. Installation 3.3 HyperCOOL Connection [HyperCOOL as a 'Cold Trap'] ① ⑧ ⑦ ② ⑥ ③ ④ ⑤ ① Rubber Valve Closed ② Tube Connection to Vacuum Pump ③ Mains Inlet ④ Vacuum Pump to HyperCOOL ⑤ Vacuum Pump to Power Outlet ⑥...
Page 13
3. Installation 3.3 HyperCOOL Connection [HyperCOOL (HC-CPP) as a ‘Cold Trap’] ⑥ ⑤ ① ④ ② ③ ① Tube Connection to Vacuum Pump ② Mains Inlet ③ Vacuum Pump to HyperCOOL ④ Vacuum Pump to Power Outlet ⑤ Mains Inlet ⑥...
4. Operation 4.1 Key Functions of Control Panel [KEYLOCK] [TIME] DEFROST, VACUUM or TIME function can be LED On: TIME setting mode locked during operation. ▶ Set H:M using the control buttons. ▶ Locked: LED On (←, ▲, ▼, SET) [DEFROST] [Display] Chamber defrosting by hot gas injection.
4. Operation 4.2 Display Temperature: displays actual temperature. (℃) Time: displays operating time (H:M) ▶ The range of time displayed is 999 hours and 59 minutes. 4.3 Defrost Defrost function warms the wall of the condenser and facilitates the release and easy removal of the “ice-plug”. Defrosting function is activated with the LED on when the DEFROST button is pressed.
4. Operation 4.4 Vacuum VACUUM function is to turn on or off the vacuum pump, where the VACUUL LED indicates current status of the vacuum pump. ▶ While the vacuum pump is resting, the VACUUM LED flashes. ▶ While the vacuum pump is running, the VACUUM LED is ON. The vacuum pump is turned off by pressing VACUUM button.
4. Operation 4.5 Time Operating time can be set up to 999 hours and 59 minutes, or ‘continuous running’ is also available. 1. Press the TIME button. 2. Press the DEFROST(up) or VACUUM (down) button to up or down the value. 3.
4. Operation 4.6 Keylock The KEYLOCK function is available to prevent unwanted interruption of operation or change of the current setting. 1. The Keylock mode is activated when the KEYLOCK button is pressed. 2. The Keylock mode is deactivated by pressing KEYLOCK once again.
5. Maintenance 5.1 Outer Part of the Device 1. Clean the outside of the device with dry soft cloth. If necessary, dip the cloth in neutral detergent and clean any contaminants. Keep it completely dry after cleaning. 2. Do not use any volatile chemicals such as alcohol, benzene, benzole, and thinner, etc. 3.
5. Maintenance 5.4 Fuse Replacement Fuse replacement may be required when the main power does not turn on by switching on the power. Please, follow the below instruction. 1. Remove the AC power cord and open the fuse case using a (-) screw driver. 2.
6. Troubleshooting 6.1 Possible Problems If any problems occur in the device, please check the following list before contact your local Hanil Scientific Inc. partner or Hanil Scientific Inc. Symptom Recommended Action - Make sure the AC Power cord completely connects the device to the Switching On mains outlet.
This message may come up if the temperature sensor of the Error 7 temperature chamber does not normally work. sensor Turn off the main power and call Hanil service. ▶ Before contact Hanil Service, please check the serial number of the device indicated on the product label. 22/26 www.ihanil.com...
6.3 Compliance 1. Comply with the instruction on the safety label. 2. Use genuine Hanil Scientific Inc. products and parts. The best performance can be guaranteed only for the Hanil Scientific Inc.-approved products. The customer service may not be provided for using any unauthorized products.
Display and Time Control incl. a Vacuum Hose and Clamps (110V, 50/60Hz) HyperCOOL Cover Plate HyperCOOL Trap Plate for Connection Vacuum Hose to Vacuum HHC-CPP Concentrator HyperCOOL HC3055, HC3110 Trap Plate for Connection Vacuum Hose HHC-CPP-A to Vacuum Concentrator with Table A(GVCT-MA) HHC-CPB HyperCOOL Acrylic Base for Manifold or Chamber...
Page 25
7. Ordering Information Rack, Flask, and Others Stainless Steel Rack with 3 Sets of SS Shelves and Trays, ø25cm HHC-CR25 (Trays can be inserted up to 5) HHC-CR-TS A Set of a Tray and a Shelve FD2009 Freeze Drying Glass Flask 40 mL (Ø45 mm) FD2010 Freeze Drying Glass Flask 80 mL (Ø...
Page 26
7. Ordering Information HyperCOOL PACKAGE INFORMATION Manifold Type Freeze Drying System -55°C with 4 Valves Basic incl. a Rotary Vane Pump, Oil Mist Trap & Connections HHC-3055-M4B (HHC3055 + HHC-CPB + HHC-MFB-4V + HHC-VH + HVP-W2V20 + HVP- WOF150) Manifold Type Freeze Drying System -110°C with 4 Valves Basic incl. a Rotary Vane Pump, Oil Mist Trap &...