Diagnostic commands

The PTU200 series digital transmitters respond to a set of diagnostic
commands, which are useful for analyzing the basic settings and
general operation of the transmitters. The diagnostic commands are
listed in
T

? Basic information on the transmitter settings

The ? command lists the basic information of the transmitter.
>?
Software version
Serial number
Configuration
Linear adjustments
Multipoint adjustments
Calibration date
Baud Parity Data Stop Dpx 9600
Echo
Sending mode
Address
Output interval
Output format
Error output format
SCOM format
Pressure unit
Temperature unit
Transducer 1:
TYPE: PMT
Transducer 2:
TYPE: VMT
Transducer 3:
TYPE: NONE
>
NOTE
Use this command only for one transmitter at a time. Any PTU200
series transmitter will always respond to ?-command whatever its
settings are, provided that the serial interface settings are correct.
44 ___________________________________________________________________ M210195
All manuals and user guides at all-guides.com
Diagnostic commands
7-2
ABLE
Function
basic information on settings
software version
serial number
outputting error messages
? <cr>
PTU200 / 1.02
T5030004
12
ON
ON
1999-03-12
ON
STOP
0
3 s
4.1 P " " UUU " " 2.1 RH " " UUU #r #n
hPa
'C
Command
?
VERS
SNUM
ERRS
E 7 1 F
-A
EN