Chapter 3

3.0 Introduction

The SQM-160 has two slots to accommodate option cards. The Option A slot holds a
Four Sensor Card, which expands the SQM-160 inputs to a total of six quartz crystal
sensors. The Option B Analog Output Card provides Rate and Thickness outputs for
connection to strip chart recorders.

3.1 Installation

W
: Hazardous voltages are present inside the SQM-160 when it is
ARNING
connected to a power source. Remove the power cable before opening the case
to install option cards.
Option A Card
4
3
Sensor 1 Sensor 2
Manufactured By
S i g m a
i n s t r u m e n t s
Remove
Top Cover.
Remove Option
Cover Plate
Insert
Option Card
Secure
Option Card
Replace Top
Cover
Option B Card
Rate
5
6
RS-232
Model No. SQM-160
Ser No. 80901
Use a 1/16" hex driver to remove the four screws that attach
the top cover to the chassis. Remove the top cover.
Use a 1/16" hex driver to remove the two 4-40 back panel
screws that hold the option cover plate. Save the screws.
Remove the option cover plate.
Insert the option card, from the inside, through the rear panel
opening. Option A slot is for the Four Sensor Card, Option B
slot is the Analog Output Card. Align the 36-pin option card
connector with the mating pins on the main board. Press the
option card connector onto the main board pins.
Secure the option card to the rear panel with the screws saved
in Step 2, and to the main board with the supplied 4-40 screw.
Replace the top cover, and attach it to the chassis with the four
screws from Step 1.
Ground
Thick
Relay I/O
Line Fuse
Fuse 1/2 A SB at 120VAC
Fuse 1/4 A SB at 240VAC
Options