Replacement of fuse

CAUTION
Be sure to find the cause of the blowout and take necessary measures before replacing the fuse.
(1) Turn off the power switch.
(2) The fuse is located at the lower left viewed from the front. Pull up the cap of the fuse holder
while rotating it counterclockwise, and the cap and the fuse are detached.
(3) Replace the fuse with a new one (Fuse: 250V AC / 3.15A delay type).
(4) Then fasten the cap of the fuse holder.
OXYGEN ANALYZER
TEMP.
T3.15A
Fuse holder (Cap)
6.4

Output characteristic

Table 1 ZX8D output characteristic
OUTPUT (mV)
%O
2
0.01
168.57
0.02
153.3
0.03
144.36
0.04
138.02
0.05
133.1
0.1
117.83
0.2
102.56
0.3
93.62
0.4
87.28
0.5
82.36
0.6
78.35
0.7
74.95
0.8
72.01
0.9
69.41
1.0
67.09
2.0
51.82
3.0
42.88
4.0
36.54
5.0
31.62
6.0
27.61
E (OUTPUT: mV) = 50.74 log 21/O
1
2
3
4
5
6
AC250V
L
N
ALM
OUT
AC
INPUT
OUTPUT (mV)
%O
2
7.0
8.0
9.0
10.0
11.0
12.0
13.0
14.0
15.0
16.0
17.0
18.0
19.0
20.0
20.6
21.0
22.0
23.0
24.0
25.0
%
2
7
F0602.ai
24.21
21.27
18.67
16.35
14.25
12.33
10.57
8.93
7.41
5.99
4.66
3.40
2.21
1.08
0.4238
0
-1.025
-2.005
-2.943
-3.84
IM 11G00E01-01EN