1. Features

2. Specifications

3. Method of measurement

4. Maintenance
5. Calibration
1. Features
The hardness of plastics is most commonly measured by
the Shore Durometer, using either the Shore A or Shore D
scale. It is the preferred method for measuring rubbers/
elastomers and it's also used for "softer" plastics such as
polyolefins, fluoropolymers and vinyls.
Shore A scale is used for "softer" rubbers, while Shore D
scale is used for "harder" ones.
Shore C/ Shore 0 scale is commonly used for tests with
foam rubbers, sponges or microporous plastics.
Designed in regard to these standards:
DIN 53505
ASTM D2240
ISO 868
HB&TI-BA-e-1311
Sauter GmbH
Ziegelei 1
D-72336 Balingen
E-Mail: info@sauter.eu
HB & TI
2. Specifications
Model
Indenter
Tip
dimension
Depth of in
dentation
Test
pressure
Measuring
spring
force
Display
range
Scale
Diameter
Weight.net
(gross)
Dimen-
sions
Thread
3. Method of measurement
Shore Durometer, like many other hardness testers,
measures the depth of indentation in the material created
by a given force on a standardized presser foot. The depth
is dependant on the hardness of the material, its
viscoelastic properties, the shape of the presser foot as
well as the duration of the test. Shore durometers allow the
measurement of the initial hardness or the indentation
hardness after a given period of time.
The basic test requires applying the force in a constant
manner without shocks, measuring the hardness (depth of
the indentation). If a timed hardness is desired, force is
applied for the required time and then the measurement
result is to be read.
The tested material should have a thickness of minimum
6.4 mm (0.25 inch).
4. Maintenance
After testing the instrument is to be put back in its packing
box. It should not be stored in following environmental
conditions: wet or dusty area, oil or chemicals.
5. Calibration
First the test plate is to be put on a hard and flat base.
The hardness tester is to be located on the test plate by
placing the measurement tip into the hole of the test plate.
The instrument is to be adjusted by turning the external
ring of the round Display to the setpoint of the test plate.
Tel: +49-[0]7433- 9933-199
Fax: +49-[0]7433-9933-149
Internet: www.kern-sohn.com
Shore A
Shore C/ 0
HBA 100-0 HBC100-0
Konus 35°
Durchm.1,3
SR2,5mm
0 – 2,5 mm
0 – 2,5 mm
ca.12,5 N
12,5 N
0,55–8,065N 0,55–8,065N 0,55-44,5N
Skala von
Skala von
0 – 100
0 – 100
55 mm
55 mm
250g
230g
(300g)
(350g)
26x62x115
26x62x115
(LxBxH) mm
(LxBxH) mm
M7 x 0.5
M7 x 0.5
Shore D
HBD100-0
Konus 30°
0 – 2,5 mm
50 N
Skala von
0 – 100
55 mm
250g
(300g)
26x62x115
(LxBxH) mm
M7 x 0.5
1

Advertisement

loading

Summary of Contents for sauter HBA 100-0

  • Page 1 E-Mail: info@sauter.eu Internet: www.kern-sohn.com Instruction Manual HB & TI 2. Specifications MANUAL SHORE DUROMETER Shore A Shore C/ 0 Shore D Model HBA 100-0 HBC100-0 HBD100-0 Indenter Konus 35° Konus 30° Durchm.1,3 SR2,5mm dimension Depth of in 0 – 2,5 mm 0 –...
  • Page 2 The testing material has to be put on the glass stage, the operation handle has to be moved down by force of the Thank you for buying a SAUTER Durometer Test Stand. designation weight. When the durometer touches the test The ruggedness of this stand will allow many years of use material completely, the value appears on the dial.

This manual is also suitable for: