Table A-2
Standard Class Assignments for the PNA Series Network Analyzer
X, P, and K 11644A
SOLT Class
Assignments
Calibration Kit Name:
X11644A, P11644A,
K11644A
Class
S
A
4
11
S
B
5
11
S
C
1, 2
11
Forward
6, 7
b
transmission
S
A
4
22
S
B
5
22
S
C
1, 2
22
Isolation
2
a. Select the LRL line auto characterization checkbox to
compute the propagation characteristics of the thru
standard from the computed propagation
characteristics of the line standard and the defined
delays for both the thru and line standards. Note, LRL
line auto characterization will only be used at
frequencies where both the THRU and LINE
standards are delay lines and where the offset
impedance of the THRU and LINE standards are
equal.
b. Forward match, reverse match, and reverse
transmission usually use the same class assignments
as forward transmission. You can select these same
class assignments in the PNA calibration menu titled
"Modify SOLT Calibration Class Assignments." Select
the checkbox titled "Link FWD TRANS, FWD
MATCH, REV TRANS and REV MATCH."
Standard Class Assignments
TRL Class
a
Assignments
Calibration Kit Name:
X11644A, P11644A,
K11644A
Class
TRL thru
6
TRL reflect
4
TRL
7, 2
line/match
Isolation
2
_
_
_
_
Cal Ref Z0 = Line Z0
Transport Ref Plane =
Thru Standard
Standard Definitions
A-5

Advertisement