Table 5-4 Using a Full 2-Port Calibration X11644A WR-90 (X-band)
Test
S
, shorted
11
S
, shorted
22
S
, thru
11
S
, thru
22
S
, thru
21
S
, thru
12
Table 5-5 Using a Full 2-Port Calibration P11644A WR-62 (P-band)
Test
S
, shorted
11
S
, shorted
22
S
, thru
11
S
, thru
22
S
, thru
21
S
, thru
12
Table 5-6 Using a Full 2-Port Calibration K11644A WR-42 (K-band)
Test
S
, shorted
11
S
, shorted
22
S
, thru
11
S
, thru
22
S
, thru
21
S
, thru
12
X, P, and K 11644A
Measured
Acceptable
Value
Value
≤ 0.15 dB
≤ 0.15 dB
≤ −40 dB
≤ −40 dB
≤ 0.06 dB
≤ 0.06 dB
Measured
Acceptable
Value
Value
≤ 0.26 dB
≤ 0.26 dB
≤ −37 dB
≤ −37 dB
≤ 0.07 dB
≤ 0.07 dB
Measured
Acceptable
Value
Value
≤ 0.40 dB
≤ 0.40 dB
≤ −35 dB
≤ −35 dB
≤ 0.01 dB
≤ 0.01 dB
Pass
(Yes/No)
p-p
p-p
p-p
p-p
Pass
(Yes/No)
p-p
p-p
p-p
p-p
Pass
(Yes/No)
p-p
p-p
p-p
p-p