Agilent Technologies 11644A Series User's And Service Manual page 45

Waveguide calibration kits, serial number prefix 3032a
Table of Contents

Advertisement

Table 5-1 Using a TRL Calibration X11644A WR-90 (X-band)
Test
S
, shorted
11
S
, shorted
22
S
, thru
11
S
, thru
22
S
, thru
21
S
, thru
12
Table 5-2 Using a TRL Calibration P11644A WR-62 (P-band)
Test
S
, shorted
11
S
, shorted
22
S
, thru
11
S
, thru
22
S
, thru
21
S
, thru
12
Table 5-3 Using a TRL Calibration K11644A WR-42 (K-band)
Test
S
, shorted
11
S
, shorted
22
S
, thru
11
S
, thru
22
S
, thru
21
S
, thru
12
X, P, and K 11644A
Measured
Acceptable
Value
Value
≤ 0.15 dB
≤ 0.15 dB
≤ −45 dB
≤ −45 dB
≤ 0.017 dB
≤ 0.017 dB
Measured
Acceptable
Value
Value
≤ 0.26 dB
≤ 0.26 dB
≤ −42 dB
≤ −42 dB
≤ 0.021 dB
≤ 0.021 dB
Measured
Acceptable
Value
Value
≤ 0.40 dB
≤ 0.40 dB
≤ −40 dB
≤ −40 dB
≤ 0.030 dB
≤ 0.030 dB
Pass
(Yes/No)
p-p
p-p
p-p
p-p
Pass
(Yes/No)
p-p
p-p
p-p
p-p
Pass
(Yes/No)
p-p
p-p
p-p
p-p
Performance Verification
System Operation Checks
5-11

Advertisement

Table of Contents
loading

This manual is also suitable for:

11644x11644p11644k

Table of Contents