Use, Maintenance, and Care of the Devices
Cleaning the Mating Plane Surfaces
These electrostatic effects can damage the device. Refer to
earlier in this chapter for additional information.
WARNING
Keep isopropyl alcohol away from heat, sparks, and flame. Store in a
tightly closed container. It is extremely flammable. In case of fire, use
alcohol foam, dry chemical, or carbon dioxide; water may be
ineffective.
Use isopropyl alcohol with adequate ventilation and avoid contact
with eyes, skin, and clothing. It causes skin irritation, may cause eye
damage, and is harmful if swallowed or inhaled. It may be harmful if
absorbed through the skin. Wash thoroughly after handling.
In case of spill, soak up with sand or earth. Flush spill area with
water.
Dispose of isopropyl alcohol in accordance with all applicable
federal, state, and local environmental regulations.
2. Clean the Mating Plane Surfaces
a. Apply a small amount of isopropyl alcohol to a lint-free cleaning swab.
b. Clean the mating plane surfaces.
c. Let the alcohol evaporate, then blow the mating plane surface dry with a gentle
stream of clean, low-pressure compressed air or nitrogen. Always completely dry a
device before you reassemble or use it.
3. Inspect
a. Inspect the mating plane surface to make sure that no particles or residue remain.
"Visual Inspection" on page
4-4
4-3.
"Electrostatic Discharge"
X, P, and K 11644A

Advertisement

Table of Contents
loading