+GF+ Signet 4630 Operating Instructions Manual page 47

Free chlorine/chlorine dioxide analyzer system
Table of Contents

Advertisement

Specifi cations
2724 DryLoc pH Electrode
General
Compatibility:
Signet 2750 pH/ORP Sensor electronics
Process
Connection:
¾ in. NPT or ISO 7/1 R¾ threads
or Signet fl ow fi ttings
Wetted Materials: PPS, glass, UHMW PE, FPM
Performance
• Effi ciency:
>97% @ 25 °C (77 °F)
Effi ciency indicates the "wellness" of a
new electrode. Effi ciency is measured by
comparing the actual slope (mV/pH) at 25
°C to the theoretical output of 59.16 mV/pH.
An effi ciency of 97% to 100% is equivalent
to a slope of 57.39 to 59.16 mV/pH.
• Range:
0 to 14 pH
Environmental
Pressure
• Operating:
-10 ºC to 65 °C (14 °F to 149 °F),0 to 6.9 bar (0 to 100 psi)
65 °C to 85 °C (149 °F to 185 °F),
Linearity derated 6.9 to 4.0 bar (100 to 58 psi)
Temperature
• Storage:
0 °C to 50 °C (32 °F to 120 °F).
CAUTION: The electrode glass will
be shattered if shipped or stored at
temperatures below 0 °C (32 °F).
The performance life of the electrode will
be shortened if stored at temperatures
above 50 °C (120 °F).
• Operating:
-10 ºC to 85 ºC (14 ºF to 185 ºF)
Rel. Humidity:
0 to 95% indoor/outdoor non-condensing to
rated ambient
Shipping Weight: 0.25 kg (0.55 lb)
Standards and Approvals
• WEEE
• RoHS Compliant
• Manufactured under ISO 9001 for Quality,
ISO 14001 for Environmental Management and
OHSAS 18001 for occupational health and safety.
China RoHS (Go to www.gfsignet.com for details)
463X Chlorine Systems Manual
47