+GF+ Signet 4630 Operating Instructions Manual page 21

Free chlorine/chlorine dioxide analyzer system
Table of Contents

Advertisement

8630 Options Menu
Test Relay 1:
>
Test Relay 2:
>
Read Sens Data:
>
No
If Yes was chosen:
Cl Sensor S/N:
xxxxxxxxx
Cl Type & Range:
2630
xxx.x ppm
Zero Cal: ppm&nA
xxx.xx
xxxx.x
In Proc: ppm&nA
xxx.xx
xxxx.x
Temp at Cal:
xxxx.x °C
pH at Cal:
xxx.xx pH
Temp Offset:
xxxx.x °C
Elapsed Time:
xxxxx.
hrs
Low & High: °C
-xxxx.x
+xxxx.x
Press UP or DOWN keys to manually toggle Relay 1 Off and On. The left LED on the front of
the transmitter confi rms operation.
Press UP or DOWN keys to manually toggle Relay 2 Off and On. The right LED on the front of the
transmitter confi rms operation.
If "YES" is selected the following (Read Only) screens will be shown.
If "NO" then this menu ends the Options Menu.
View the sensor serial number.
Identify the chlorine sensor type connected and its ppm range.
View user entered Zero Calibration data in ppm and nA.
View user In-Process Calibration value when it was entered in the Calibrate Menu.
Temperature recorded during user In-Process Calibration.
pH value recorded during user In-Process Calibration.
Temperature offset calculated from user entered temperature calibration from Calibrate Menu.
Total hours of operation.
Lowest and highest temperatures the Cl sensor has been subjected to during operation.
End of Options Menu
463X Chlorine Systems Manual
21

Advertisement

Table of Contents
loading

This manual is also suitable for:

Signet 4632

Table of Contents