+GF+ Signet 4630 Operating Instructions Manual page 19

Free chlorine/chlorine dioxide analyzer system
Table of Contents

Advertisement

8630 Calibrate Menu
If Window Mode was chosen:
Relay 2 Source:
>
pH
Relay 2 Rng: ppm
>
0.00
5.00
Relay 2 Hys:
>
0.20
ppm
Relay 2 Delay:
>
0.0
secs
If Pulse Mode was chosen:
Relay 2 Source:
pH
Relay2 Rng: pH
0.00
14.0
Relay2 PlsRate:
120 pulses/min
Find New
>
Sensor(s):
Last Cal
>
10-18-09
Choose pH or chlorine for Relay 2.
Enter the range where Relay 2 will activate above and below this setpoint.
This relay will be deactivated at Relay 2 Range setpoints ± this hysteresis.
Set the time delay for Relay 2 to activate after reaching the setpoints set in the
Relay 2 Range.
Choose pH or chlorine for Relay 2.
>
>
Enter the range where Relay 2 will activate above and below this setpoint.
Set the maximum pulse rate. Range: 1–400 pulses/min.
>
Required only when a new sensor is changed while the power is on. Choose Yes or No.
Edit the calibration date.
End of Calibrate Menu
463X Chlorine Systems Manual