Micro-Ohmmeter Model 5600
Battery Life (Typical):
Ω
2000µ
range: 200 to 300 8-second measurements
Ω
20-200m
ranges: 900 30-second measurements (approx)
Ω
Ω
2000m
, 20-200
ranges: 5000 30-second measurements (approx)
Operating Voltage: 4.5 to 7V
DC
Recharge Time: 14 hours from full discharge (approx)

Mechanical Specifications

Display: 2000ct, 7 segment LED
Case: Fiberglass-charged polycarbonate, with handle and removable lid,
watertight to IP534 (lid closed), IP520 (lid open).
Color: Safety yellow case with gray faceplate
Operating Temperature: 14° to 131°F (-10° to 55°C)
Storage Temperature: -4° to 131°F (-20°C to 55°C)
Dimensions: 15.35 x 10.24 x 9.84" (390 x 260 x 250mm)
Weight: 17 lbs 10 oz (8kg approx) without leads

Safety Specifications

Safety Standards:
Electrical: EN 61010, 300V, Cat. III
Electromagnetic:
IEC 801-2, Electrostatic discharge
IEC 801-3, Electromagnetic Fields
IEC 801-5, Electrical shocks
Mechanical/environmental:
IEC 529, Protection Index (IP)
IEC 68.2.6, Vibrations
IEC 68.2.27, Shocks
IEC 68.2.29, Shake
IEC 68.2.31, Knock over
IEC 68.2.32, Drop test
Case Material: UL 94
Overload Input Protection: 380Vrms
Fuses: Two high interrupting capacity, 8A, 380V, 1¼ x ¼"
- 6 -