Example Settings; Emc Compliance; Warranty - Young 86004 Instructions Manual

Ultrasonic anemometer
Table of Contents

Advertisement

6.4.13 SET13n ASCII and ANALOG OUTPUT FORMAT
This setting determines whether ASCII and ASCII POLLED serial
outputs and analog outputs provide wind data in either Polar (speed
and direction) or Cartesian (UV) form.
6.4.14 SET14n ANALOG OUTPUT STATUS
Serial ASCII and ASCII POLLED formats report a STATUS CODE
where non-zero values indicate insufficient samples or measurement
error. SET14n determines how the STATUS CODE is indicated by
the analog output.
6.4.15 SET15n ASCII RESOLUTION
Sets low or high wind speed and direction resolution.
6.4.16 SET16n FORCE ANALOG OUTPUT
This command forces both analog output channels to LO, MID, or
HI scale. This may be used to calibrate or check the operation of
externally connected analog measurement devices.
6.4.17 SET20n HEATERS
This command enables or disables the 86004 heaters. Heaters are
enabled by default.
6.4.18 SET77 SAVE SETTINGS
Use this command to save current parameter settings to non-volatile
memory. Any settings which have not been saved will be lost when
power is removed. Saved settings are loaded at power up. This
command may be used any time the sensor is in COMMAND MODE.
6.4.19 XX, RPTV, and ??
XX
Returns the sensor to OPERATE MODE.
RPT
Reports the current parameter settings.
??
Shows a list of commands.

7.0 EXAMPLE SETTINGS

Suggested settings. Not all possible setting combinations are shown.
Using YOUNG 86SETUP program for changing parameters is
recommended. See wiring diagrams for jumper settings.
7.1 FACTORY DEFAULT
RS-232
Enabled
RS-485 Output Only Enabled
Voltage Output:
Enabled
Current Output:
Disabled
Serial Output Format: RMYT
Output Interval
500
Sample Count:
16
7.2 4-20 mA OUTPUT
RS-232
Enabled
RS-485
Disabled
Voltage Output:
Disabled
Current Output:
Enabled
Output Interval
1000
Sample Count:
50
7.3 RS-485 POLLED ASCII OUTPUT
RS-232
Enabled
RS-485 (Full Duplex) Enabled
Voltage Output:
Disabled
Current Output:
Disabled
Serial Output Format: ASCII POLLED
Sample Count:
50
7.4 HIGH WINDS
Output Delay:
1000
Sample Count:
200

8.0 EMC COMPLIANCE

This sensor complies with limits for a Class A digital device, pursuant
to part 15 of the FCC Rules, and IEC standard 61326-1. This sensor
generates, uses, and can radiate radio frequency energy and, if
not installed and used in accordance with the instruction manual,
may cause harmful interference to radio communications. Sensor
operation may be temporarily affected by radio frequency and
transient interference sources, but will revert to proper operation
when the source of interference is removed.

9.0 WARRANTY

This product is warranted to be free of defects in materials and
construction for a period of 12 months from date of initial purchase.
Liability is limited to repair or replacement of the defective item. A copy
of the warranty policy may be obtained from R. M. Young Company.
Page 5
86004-90(G)

Advertisement