Fluke 87 V/AN Service Manual page 40

Table of Contents

Advertisement

87 V/AN
Service Manual
Set the calibrator to Standby prior to changing the function switch position
and or after completing adjustment of each function.
If the calibration adjustment procedure is not completed correctly, the
Meter will not operate correctly.
Switch Position
(Function)
K
(AC Volts)
L
(DC Volts)
m L
(DC Millivolts)
(Ohms)
O
(Diode Test)
A, mA
(Amps, milliamps)
UA
(Microamps)
32
Table 16. Calibration Adjustment Steps
Input
Calibration Adjustment
Terminal
C-01
I
C-02
C-03
C-04
C-05
C-06
C-07
C-08
C-09
C-10
C-11
C-12
C-13
C-14
C-15
C-16
C-17
C-18
C-19
C-20
C-21
C-22
A
C-23
C-24
mA /UA
C-25
C-26
C-27
C-28
mA /UA
C-29
C-30
C-31
Notes
Step
600.0 mV, 60 Hz
600.0 mV, 20 kHz
6.000 V, 60 Hz
6.000 V, 20 kHz
60.00 V, 60 Hz
60.00 V, 20 kHz
600.0 V, 60 Hz
600.0 V, 10 kHz
6.000 V, 0 Hz
60.00 V, 0 Hz
600.0 V, 0 Hz
600.0 mV, 0 Hz
60.00 mV, 0 Hz
600.0 Ω
6.000 kΩ
60.00 kΩ
600.0 kΩ
6.000 MΩ
0.000 Ω
50.0 MΩ
3.000 V, 0 Hz
6.000 A, 60 Hz
6.000 A , 0 Hz
60.00 mA, 60 Hz
400.0 mA, 60 Hz
60.00 mA, 0 Hz
400.0 mA, 0 Hz
600.0 µA, 60 Hz
6000 µA, 60 Hz
600.0 µA, 0 Hz
6000 µA, 0 Hz
Input Value