Reassembling The Meter Case; Replacing The Battery; Testing Fuses And Current Circuitry - Fluke 87 V/AN Service Manual

Table of Contents

Advertisement

87 V/AN
Service Manual

Reassembling the Meter Case

To reassemble the Meter case:
1. Verify that the rotary knob and circuit board switch are in the
2. Place the PCA into the bottom case.
3. Place the case top on the case bottom.
4. To avoid damaging the battery wire, ensure the wire exits the middle of the battery
5. Properly seat the case gasket and snap the case halves together above the LCD end.
6. Reinstall the three case screws and the battery door.
7. Secure the battery door by turning the screw 1/4-turn clockwise.
8. Go to "Performance Tests" later in this document, and perform the procedures

Replacing the Battery

Replace the battery with a 9-V battery (NEDA A1604, 6F22, or 006P).
To avoid false readings, which could lead to possible electric
shock or personal injury, replace the battery as soon as the
battery indicator (b) appears. If the display shows "bAtt" the
Meter will not function until the battery is replaced.
Replace the battery as follows, refer to Figure 2:
1. Turn the rotary knob to OFF and remove the test leads from the terminals.
2. Remove the battery door by using a standard-blade screwdriver to turn the battery
3. Remove the old battery and replace it with a new one.
4. Align the battery leads so that they not pinched between the battery door and the
5. Secure the door by turning the screws one-quarter turn clockwise.

Testing Fuses and Current Circuitry

If a test lead is plugged into the mA/µA or A terminal and the rotary knob is turned to a
non-current function, the Meter chirps and flashes "LEAd" if the fuse associated with
that current terminal is good. If the Meter does not chirp or flash "LEAd", the fuse is bad
and must be replaced. Refer to Table 17 for the appropriate replacement fuse.
Before measuring current, test the quality of the appropriate fuse and the current shunt
using the following procedure. See Figure 4.
1. Turn the rotary knob to N.
2. To test F2, insert a test lead into the I input terminal and touch the probe to
22
the gasket remains secured to the bottom case.
compartment.
See Figure 2.
described.
door screws one-quarter turn counterclockwise.
case bottom.
the A input terminal.
The input receptacles contain split contacts. Be sure to touch the probe to
the half of the receptacle nearest the LCD.
XWWarning
Note
OFF
position, and that

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents