BEFORE USING ANY ANALYSER CHECK THAT:
- The calibration date has not expired
- The particle filter is not dirty
- The water trap and flue probe hose are empty of water
- The water trap is fitted correctly to the instrument including the
drain plug
- The flue probe hose is connected properly to the water trap
- The flue temperature plug is connected into Tf
USING THE ANALYSER
Switch the analyser ON
It will now purge for approximately 20 seconds after which it will show a
set a readings on the display and now is a good time to check the
battery status.
Insert the probe into the centre of the flue after approximately 60
seconds the readings should stabilise.
With the probe in the flue you can now print a live test result. Switch on
your printer and line up the bottom of it, with the top of your analyser,
now press the PRINTER key
If you want to store the reading in the analyser memory, again with the
probe in the flue, hold down the STORE key
shows you the test number at which it will log your readings.
To retrieve a stored reading dial the rotary switch down to the menu
position and using either of the arrow keys ↑ or ↓ scroll through the
menu to the store feature. Press the STORE key
view, select the test number you want to print by using the arrow keys to
count up (press lightly until you hear a beep). Once you have selected
your results, press the PRINT key
The analyser will automatically purge for 30 seconds when switched
OFF
.
.
until the display shows printing.
.
until the display
and scroll to