CLEANING, DECONTAMINATION AND STERILIZATION
Validated manual cleaning and disinfection procedure
STEP STEP DESCRIPTION STEP INSTRUCTION
Contamination
Cleaning
1
Removal
Steps
Rinse
2
Drying
3
Disinfection
Disinfection
4
Application
Steps
Manual
5
Disinfection
Final Rinse
6
Final Drying
7
WARNING: Zimmer® UNIVERSAL Power System should not be immersed or submerged in water or cleaner.

PREPARATION OF AUTOMATED CLEANING MATERIALS

The automated cleaning and disinfection process has been validated according to AAMI TIR Number 30 and ISO 17664 guidelines.
CAUTION: The manual cleaning and disinfection procedure must be performed prior to the automated cleaning and
disinfection procedure.
Materials:
Detergent with neutral pH or lightly alkaline: pH < 9
Neutralizing agent
Rinse aid
Automated washer
The UNIVERSAL lift out tray (REF 89-8510-459-48) and the UNIVERSAL sterilization case (REF 89-8510-459-47) are also sold as
a kit (REF 89-8510-459-46). The use of the lift out tray (REF 89-8510-459-48) has been validated and is recommended for the
automated cleaning and disinfection procedure following the manual cleaning and disinfection procedure.
ACCESSORIES
− Cold/Room temperature
Rinse product with RO or
distilled water removing
RO or distilled water
any visible organic
− Soft bristle brush
− Non-shedding cloth
material. Apply detergent
− Recommended pH-value
with a non-shedding cloth
detergent/ disinfectant
and as necessary, with
assistance of a soft bristle
to brush any visible soil.
− Cold/Room temperature
Rinse product with RO or
distilled water.
RO or distilled water
Dry the device utilizing a
− Non-shedding cloth
− Medical quality filtered
dry non-shedding cloth.
Medical quality filtered
compressed air
air may be utilized if
available.
− Recommended pH-value
Apply the recommended
pH-value detergent/
detergent/ disinfectant
− Non-shedding cloth.
disinfectant
− Spray bottle or other
to a non-shedding cloth.
Cover the device's surface
manual applicator
area with the saturated
cloth.
After 30 minutes contact
− Recommended pH-
time, wipe the device
value detergent/
contact surfaces, joints
disinfectant
and mated areas utilizing
− Non-shedding cloth
the cover cloth.
− Room temperature RO
Rinse product under
room temperature RO or
or distilled water
distilled water.
Dry the device utilizing a
− Non-shedding cloth
− Medical quality filtered
dry non-shedding cloth.
Medical quality filtered
compressed air
air may be utilized if
available.
16
DURATION
Until all visible soil is
removed.
Minimum of two (2)
minutes is recommended.
Until all visible detergent
is removed.
Until product is visually
dry.
Minimum of thirty
(30) minutes is
recommended.
Manual disinfection
is complete when the
device's surface, joints
and crevices have been
manually wiped.
Minimum of 30 seconds.
Until product is visually
dry.