INDICATIONS FOR USE

The VariSoft
cannula and tubing make up the infusion set.

CONTRAINDICATIONS

VariSoft is neither intended nor indicated for use with blood or blood products (I.V. infusion).

WARNINGS

• VariSoft infusion sets with t:lock tubing connectors should only be used with Tandem cartridges featuring the t:lock
connector.
• The VariSoft infusion set is only sterile and non pyrogenic if the packaging is unopened and undamaged. Do not use it if the
packaging is already opened or is damaged.
• Go through instructions for use carefully before inserting VariSoft, as failure to follow instructions may result in pain or injury.
• Inaccurate medication delivery, infection and/or site irritation may result from improper insertion and maintenance of
infusion site.
• When using VariSoft for the first time, do so in the presence of a healthcare provider.
• VariSoft is a single-use item, which must be disposed of after usage. Do not clean or re-sterilize.
• Be sure that the needle guard is removed before insertion.
• People with small or large amounts of subcutaneous fat should choose the insertion angle carefully, as the cannula could
be placed in underlying muscle or dermal layer and hereby cause reduced or blocked medication delivery. Normal insertion
angle between 20-45°. Consult your healthcare provider concerning this matter.
• Do not leave air in the infusion set. Make sure to fill VariSoft completely.
Instructions for Use
• Do not in any way put disinfectants, perfumes, deodorants or other products containing alcohol or disinfectants in contact
with the tubing connector and the tubing, as these may affect the integrity of the infusion set.
• Replace the infusion set if the adhesive tape becomes loose.
• Check the infusion set frequently to ensure that the soft cannula remains firmly in place, and replace if it is not secured. Since
the cannula is soft, it will not cause any pain if the soft cannula becomes fully or partially dislodged from the skin, and this
may take place without notice. The soft cannula must always be completely inserted to receive full amount of medication.
• Replace the infusion set every 48-72 hours, or per your healthcare provider's instructions.
• If infusion site becomes inflamed, replace set and use new site until the first site has healed.
• Do not re-insert introducer needle into infusion set. Re-insertion could cause tearing of soft cannula which would result in
Use by (year-month-day)
unpredictable medication flow.
• Never prime the infusion set or attempt to free a clogged tubing while set is inserted. You may accidentally inject too
Do not re-use
much medication.
• Use aseptic techniques when temporarily disconnecting VariSoft, and seal the cannula housing with the cover provided.
Consult with healthcare provider on how to compensate for missed medication when disconnected.
Sterilized using ethylene oxide
• Protect VariSoft from direct sunlight. Store at room temperature.
• Reuse of the infusion set may cause infection, site irritation, or damage to the cannula/needle. A damaged cannula/needle
Attention: See accompanying documents
may lead to inaccurate medication delivery.
• When infusing insulin, carefully monitor blood glucose levels when disconnected and after reconnecting.
See Instructions for Use
• When infusing insulin, and blood glucose level becomes unexplainably high, or occlusion alarm occurs, check for clogs and
leaks. If in doubt, change infusion set since soft cannula could be dislodged, crimped or partially clogged. Discuss plan for
Reference / model number
rapid replacement of insulin with healthcare provider should any of these problems arise. Test blood glucose level to make
sure problem is corrected.
• When infusing insulin, check the blood glucose level 1-2 hours after introducing VariSoft and check the infusion site several
Lot number
times a day, and measure your blood glucose regularly.
• When infusing insulin, do not change infusion set just prior to bedtime, unless blood glucose can be checked 1-2 hours
Manufacturer
after insertion.
Distributed by:
Legal manufacturer:
11045 Roselle Street
Unomedical a/s
San Diego, CA 92121, USA
Aaholmvej 1-3, Osted
www.tandemdiabetes.com
4320 Lejre, Denmark
+1-858-255-6269
Made in Denmark
© 2017 Tandem Diabetes Care, Inc. All rights reserved. Tandem Diabetes Care is a registered
trademark and VariSoft is a trademark of Tandem Diabetes Care, Inc. AW-1000226_A
infusion set is indicated for the subcutaneous infusion of insulin administered by an external pump. The VariSoft

CONTENTS

A.
Needle guard
G. Connector needle
B. Soft cannula
H. Circular protective cap
C. Adhesive tape
Disconnect cover
I.
D. Introducer needle
E.
Cannula housing
F.
Tubing
Fill the empty space in
the cannula with the
following prime:
U-100 insulin:
0.7 units (0.007 ml)

DISCONNECTING

Consult pump
manufacturer's
instructions for use
when connecting
infusion set to
insulin pump.

RECONNECTING

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the VariSoft and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Tandem VariSoft

  • Page 1 • VariSoft infusion sets with t:lock tubing connectors should only be used with Tandem cartridges featuring the t:lock connector. • The VariSoft infusion set is only sterile and non pyrogenic if the packaging is unopened and undamaged. Do not use it if the packaging is already opened or is damaged.
  • Page 2 • Si le dispositif VariSoft est utilisé pour la première fois, le placer en présence d’un médecin. • Le dispositif VariSoft est à usage unique et doit être jeté dans un récipient approprié après utilisation. Ne pas nettoyer ou restériliser.

Table of Contents