Tandem AutoSoft 30 Instructions For Use
Tandem AutoSoft 30 Instructions For Use

Tandem AutoSoft 30 Instructions For Use

Infusion set
Hide thumbs Also See for AutoSoft 30:

Advertisement

Available languages

Available languages

Quick Links

INDICATIONS FOR USE
The AutoSoft
DESCRIPTION
AutoSoft 30 is an angled infusion set with an integrated insertion device. The built-in legs support a 30° insertion angle. This
product is delivered ready for use. Both the inserter and the separate tubing are sterile.
CONTRAINDICATIONS
AutoSoft 30 is neither intended nor indicated for use with blood, blood products or intravenous infusion (IV).
WARNINGS
• AutoSoft 30 infusion sets with t:lock tubing connectors should only be used with Tandem cartridges featuring the t:lock
connector.
• AutoSoft 30 is only sterile and non pyrogenic if the packaging is unopened and undamaged. Do not use it if the sterile paper
or red seal has been broken.
• Go through instructions for use carefully before inserting AutoSoft 30, as failure to follow instructions may result in pain
or injury.
• Use AutoSoft 30 for the first time in the presence of your healthcare provider. When possible, stand or sit upright when
using AutoSoft 30.
• Do not put perfumes, deodorants, cosmetics or other products containing alcohol or disinfectants in contact with the
connector or the tubing. This may affect the integrity of the infusion set.
• AutoSoft 30 is a single-use device. Do not clean or re-sterilize.
Instructions for Use
• Put the lid back on the insertion device for safe disposal. For available sharps containers please consult your local pharmacy.
• Remove the needle protector before use.
• When priming, make sure there are no air bubbles in the tube.
• Inaccurate insulin delivery, infection and/or site irritation may result from improper insertion or maintenance of the
infusion site.
• Replace the infusion set every 48-72 hours, or per your healthcare provider's instructions.
• If the soft cannula bends during insertion, apply a new AutoSoft 30.
• Replace the infusion set if the adhesive tape comes loose.
• Check the infusion set frequently to ensure that the soft cannula remains firmly in place. Replace with new set, if the set is
Use by (year-month-day)
not in place. Since the cannula is soft, it will not cause any pain if it slips out, and this may happen without notice. The soft
cannula must always be completely inserted to receive the full amount of medication.
Do not re-use
• If your infusion site becomes red or inflamed, replace the infusion set and use a new site until the first site has healed.
• Do not re-insert the introducer needle into the infusion set. This could cause tearing of the soft cannula and unpredictable
medication flow.
Sterilized using ethylene oxide
• Never try to fill or free clogged tubing while the infusion set is inserted. This may result in unpredictable medication flow.
Attention: See accompanying documents
• Wash your hands before disconnecting and reconnecting the infusion set. Consult your healthcare provider on how to
compensate for missed medication while disconnected.
• Protect AutoSoft 30 from direct sunlight and humidity. Store in a dry place at room temperature.
See Instructions for Use
• Never point a loaded insertion device towards any body part, where insertion is not desired.
• Reuse of the infusion set may cause infection, site irritation, or damage to the cannula/needle. A damaged cannula/needle
Reference / model number
may lead to inaccurate medication delivery.
Lot number
RECOMMENDATIONS
• For use with insulin: Check your blood glucose level 1-2 hours after insertion to ensure correct insertion/flow.
Manufacturer
• Check the infusion site several times a day.
• Do not change your infusion set just before bedtime, unless your blood glucose can be checked 1-2 hours after insertion.
Distributed by:
Legal manufacturer:
• If your blood glucose level becomes unexplainably high or an occlusion alarm occurs, check for clogs and leaks. If in doubt,
change your infusion set.
• Carefully monitor your blood glucose levels when disconnected and after reconnecting.
11045 Roselle Street
Unomedical a/s
San Diego, CA 92121, USA
Aaholmvej 1-3, Osted
www.tandemdiabetes.com
4320 Lejre, Denmark
+1-858-255-6269
Made in Mexico
© 2017 Tandem Diabetes Care, Inc. All rights reserved. Tandem Diabetes Care is a registered
trademark and AutoSoft is a trademark of Tandem Diabetes Care, Inc. AW-1000220_A
CONTENTS
A.
B. Needle protector
C. Soft cannula
D. Adhesive
E.
F.
G. Tubing
H. Introducer needle
I.
J.
K. Lined indentations – hold here when
L.
M. Disconnect cover
N. Circular protective cap
30 infusion set is indicated for the subcutaneous infusion of insulin administered by an external pump.
Lid
Cannula housing
Connector
Insertion device
Sterile paper
loading AutoSoft 30 (see figure 7)
Release buttons – press here when
releasing AutoSoft 30 (see figure 9)
PREPARATION AND INSERTION
6.
Hold on the lined indentations with one hand and
pull the lid off with the other.
Read all instructions carefully before using AutoSoft 30.
CAUTION: Be careful not to bend or touch the
Consult your pump manufacturer's instructions for use
before connecting AutoSoft 30 to your pump. Follow
AutoSoft 30 introducer needle.
proper hygiene procedures.
IMPORTANT NOTE: If the infusion set is not
securely placed in the insertion device with the
INSTRUCTIONS
needle pointing straight ahead, pain or minor injury
1.
Wash your hands before inserting AutoSoft 30.
may occur during insertion.
2.
Recommended insertion areas. Do not use an
7.
Place your fingers on the lined indentations on
each side. Pull back the spring softly until you hear
area right next to your previous place of insertion.
a "click". The needle protector will lift up or peel off.
Please refer to your healthcare provider for infusion
Do not touch or bend the needle.
site rotation.
8.
Immediately before insertion, carefully remove the
3.
Clean the area of insertion with a disinfectant as
needle protector. Be sure the tape is not stuck on
directed by your healthcare provider. Be sure the
the introducer needle.
area is dry before inserting AutoSoft 30.
CAUTION: Check the tamper-proof seal and sterile
9.
INSERTION: Place your index finger on the top
release button and your thumb on the bottom
paper. Do not use if the seal or paper has been
release button. Position AutoSoft 30 with its
broken.
legs level (flat) on the skin to ensure an insertion
4.
Pull the red tape to remove the seal.
angle of 30°. Make sure to maintain a 30° angle
5.
Pull off the sterile paper. Do not touch the spring at
while inserting and gently squeeze once to insert
this stage.
AutoSoft 30.
DISCONNECT
A.
AutoSoft 30 allows you to temporarily disconnect from your pump. Place a finger
just in front of the cannula housing and gently squeeze the sides of AutoSoft 30.
Pull the connector needle straight out.
B. Insert the circular protective cap into the connector needle until you hear a "click".
C. Insert the disconnect cover into the cannula housing until you hear a "click".
10. Make sure that the cannula stays in place: Put your
finger gently on the clear window while removing
the insertion device and introducer needle. Pull it
gently straight back.
11. Secure the cannula with one finger on the clear
window. Carefully remove the backing paper under
the cannula housing as shown.
12. Remove the backing paper under the front end.
Massage the adhesive thoroughly onto the skin.
IMPORTANT: Do not attempt to straighten the
adhesive if curled.
13. Put the lid back on for disposal.
14. Remove the circular protective cap. Prime the
infusion set until medication dispenses from the
tip. Ensure there are no air bubbles in the tubing
or AutoSoft 30. Consult your pump manufacturer's
instructions for use for proper priming procedures.
15. Place a finger in front of the cannula housing and
insert the connector needle. Make sure the
connector locks with a "click". Fill the empty space
in the cannula with the following prime:
U-100 insulin:
0.7 units (0.007 ml)
RECONNECT
D. Remove the circular protective cap. Prime the infusion set until medication
dispenses from the needle.
CAUTION: When priming, hold AutoSoft 30 with the needle pointing down.
E.
Remove the disconnect cover from the cannula housing.
F.
Place a finger in front of the cannula housing. Push the connector needle
straight in until you hear a "click".

Advertisement

loading

Summary of Contents for Tandem AutoSoft 30

  • Page 1 Be sure the infusion set until medication dispenses from the • AutoSoft 30 is only sterile and non pyrogenic if the packaging is unopened and undamaged. Do not use it if the sterile paper the introducer needle.
  • Page 2 13. Remettre le capuchon pour l’élimination. tel que cela est indiqué par votre équipe de soins. • AutoSoft 30 est stérile et apyrogène seulement si l’emballage est fermé et intact. Ne pas utiliser si le papier stérile ou le Immédiatement avant l’insertion, retirer Vérifier que la zone est sèche avant d’insérer...