TIS CHECKBOX
INSTRUCTION MANUAL VERSION 1.00
NOTE
THE INSTRUMENT MUST ONLY BE USED BY SUITABLY
TRAINED AND COMPETENT PERSON(S).

Advertisement

loading
Need help?

Need help?

Do you have a question about the TIS CHECKBOX and is the answer not in the manual?

Questions and answers

Summary of Contents for TIS TIS CHECKBOX

  • Page 1 TIS CHECKBOX INSTRUCTION MANUAL VERSION 1.00 NOTE THE INSTRUMENT MUST ONLY BE USED BY SUITABLY TRAINED AND COMPETENT PERSON(S).
  • Page 2: Table Of Contents

    CONTENTS Introduction & Safety Checkbox Description & Maintenance Checkbox Operation Technical Specification Recommended Verification Sequence Record Form Warranty Information Optional Accessories / Test Leads: Part Number: TIS172ML Description: Universal three wire 4mm 13A Mains lead. Part Number: TIS173DB Description: Three wire distribution board lead set complete with crocodile clips and probes.
  • Page 3: Introduction & Safety

    INTRODUCTION & SAFETY TIS Checkbox is a professional, multifunctional field checkbox, intended for use with all makes and models of installation test instruments. In addition to its durable body and simple operation it boasts the following features:  Insulation calibration with test voltages up to 1000 V, ...
  • Page 4 / damage to the TIS Checkbox.  The polarity of the mains input must be verified before powering the TIS Checkbox, failure to do so may result in a safety hazard / damage to the TIS Checkbox.
  • Page 5: Checkbox Description & Maintenance

    CHECKBOX DESCRIPTION & MAINTENANCE Figure 1.0 Figure 1.1...
  • Page 6 Voltage / frequency L-PE Multimeter output. Insulation resistances outputs. LowΩ and continuity resistances outputs. △ Caution – do not use test currents higher than 300 mA. ! Incorrect test current indication for RCD trip-out time function verification (appropriate LED lights if test current is too high).
  • Page 7 (Figure 1.1, Section 17) WARNING Disconnect any connected instrument/accessory and power off the TIS Checkbox before opening fuse compartment cover. Replace blown fuse with original type only, failure to do so may cause a safety hazard and / or damage to the TIS Checkbox.
  • Page 8: Checkbox Operation

    Connect Instrument under test to the 230V 13A socket (Figure 1.0 – Section13). Check polarity of mains input & connect TIS Checkbox to mains outlet. Select Local Loop range using the RCD/LOOP button (Fig. 1 Section 8). Perform either Hi Current or NO-TRIP Impedance test.
  • Page 9 Loop Impedance Verification Loop +0.50R Select Loop+0.50R range using the RCD/LOOP button (Fig. 1 Section 8). Perform either Hi Current or NO-TRIP Impedance test. Check that the loop impedance measurement has increased from the local loop impedance reading by a nominal of 0.50R (Specification: +/-5% of Range), +/- Instrument accuracy Record measurement.
  • Page 10 RCD Test Verification (X5, 30.00mA) Select RCD 30.00mA X5 range using the RCD/LOOP button (Fig. 1 Section 8). Select & perform 30.00mA X5 RCD Test. Specification: 20.0mS +/-2mS, +/- Instrument accuracy Neither current low nor current high LED’s should illuminate whist conducting this test, however if the RCD current fails please refer to manufacturers specifications.
  • Page 11: Technical Specification

    TECHNICAL SPECIFICATION Function Range Specification Continuity Resistance Com/Null The two com/null terminals can be used to short the Customer’s leads to Null/Zero the lead resistance. Com/Null 0.500R +/- 5% Maximum input current 350mA 5.000R +/- 2% Maximum input current 350mA 10.000R +/- 2% Maximum input current 350mA Insulation Resistance...
  • Page 12: Recommended Verification Sequence Record Form

    RECOMMENDED VERIFICATION PROCESS FORM Instrument Under Test Manufacturer: Model Number: Type of Instrument: Serial Number: Traceable Information TIS Checkbox Serial Number: Date of Calibration: Certificate Number: Verification Process Function Checkbox Value / Range Measured Value: Ω Continuity >200mA Functionality 0.500Ω...
  • Page 13: Warranty Information

    NEW INSTRUMENTS HAVE A WARRANTY PERIOD OF: 1 YEAR FROM THE DATE OF PURCHASE BY THE USER. (A copy of the original purchase invoice may be requested to validate the purchase date). Any unauthorized repair / adjustment will void the warranty. For service / calibration / repair requirements, please contact: UNIT 12/14, LUDDITE WAY BUSINESS PARK RAWFOLDS WAY...

This manual is also suitable for:

Tis172mlTis173dbTis25