High Pressure Oxygen; Cooler (3K-2); Vessel - DDS Calorimeters CAL3K-AP Installation Manual

Oxygen bomb calorimeters
Hide thumbs Also See for CAL3K-AP:
Table of Contents

Advertisement

REMEMBER: You may fill the Vessel LESS than 3000KPA but NEVER more than 3000KPA.

HIGH PRESSURE OXYGEN

The type of Oxygen to be used is the standard Industrial Oxygen used for welding. A purity of 99.5% will be
suitable.
Conform to the industry standard of handling procedures for oxygen bottles.
Secure the bottle against a wall.
The unit has an exhaust outlet (6mm pipe). Route this pipe to the outside or any suitable scrubber to
prevent the burned gas to enter the room.

COOLER (3K-2)

Electrical shock: There are no user serviceable parts inside.

VESSEL

The Vessel has been pressure tested (certificate will be provided if required), at 40 MPA according to
international standards.
NEVER OPERATE ANY UNSAFE VESSEL.
NEVER IMMERSE THE VESSEL INTO COLD WATER WHEN COOLING. USE THE 3K-2 AIR COOLER OR
ALLOW THE VESSEL TO COOL DOWN NATURALLY.
A SUITABLE INSPECTION WARNING PERIOD OF 2000 FIRINGS, AND A STOP OPERATING LIMIT OF 2500
FIRINGS ARE PRE-PROGRAMMED FROM THE FACTORY
THE VESSEL MUST BE INSPECTED AT REGULAR INTERVALS DEPENDING ON THE USE. A SUITABLE INSPECTION
PERIOD CAN BE SET (By your agent!) AND THE UNIT WILL ISSUE AN INSPECTION WARNING AND SOON AFTER THIS
IT WILL SHUT-DOWN THE OPERATION.
NOTE: THE MANUFACTURER (DDS) AND ITS AGENTS CANNOT BE HELD RESPONSIBLE FOR ANY LOSS
OR DAMAGE RESULTING FROM NEGLIGENCE OR INCORRECT USE OF THEIR EQUIPMENT. IT IS YOUR
DUTY AND RESPONSIBILITY TO CHECK ALL SAFETY ASPECTS OF THE CAL3K SYSTEM REGULARLY.
CAL3K-AP INSTALLATION GUIDE V1.5 |
Manufacturers of CAL2K/CAL3K Oxygen Bomb Calorimeters
www.ddscalorimeters.com
DDS CALORIMETERS
www.ddscalorimeters.com
| calo@ddsystems.co.za

Advertisement

Table of Contents
loading

Table of Contents