5.2.1 High Pressure Oxygen; Vessel (3K-5S) - DDS Calorimeters 3K-ST Installation Manual

Table of Contents

Advertisement

NOTE: With some applications you may need to fill the Vessel to less than 3000KPA e.g.
1500KPA (15 bar), however this is only applicable when you have a sample which is highly
combustible, for example with wood related samples.
REMEMBER: You may fill the Vessel LESS than 3000KPA but NEVER more than 3000KPA.

5.2.1 HIGH PRESSURE OXYGEN

The type of Oxygen to be used is the standard Industrial Oxygen used for welding. A purity
of 99.5% will be suitable.
Conform to the industry standard of handling procedures for oxygen bottles.
Secure the bottle against a wall.
Should the Oxygen supply be in another room, make sure there is a high pressure shut off
valve near the 3K-3 Filling Station installation.
5.3
VESSEL (4K-4)
• Handle the vessel over a table or solid surface.
• Do not drop a vessel.
• Do not overfill a vessel with oxygen: The maximum oxygen filling pressure is 30bar / 3000kPA, which
is the standard for bomb calorimetry.
• Use the provided De-filler cap only, part number: 3K-3-22.
• Do not use any oil or grease on any of the oxygen fittings or the O-rings. A very small amount of
high-pressure (vacuum) grease is allowed on the O-ring to make them pliable. Wipe them clean
afterwards.
• Store the vessel with the lid open. Never store a vessel under pressure, always deflate using the de-
filler cap.
• Clean the inside of the vessel with paper towel after every use.
• Weigh the crucible every week and discard it when it has lost 10% of its original mass.
• Discard the crucible when the bottom is distorted and extremely discoloured.
• Remove the deflector plate every week and clean the top surface. It does accumulate residue and
must be clean regularly.
• Make sure to use the pipe-clamping device on the right side of the filing station for safe use.
NEVER OPERATE ANY UNSAFE VESSEL.
NEVER IMMERSE THE VESSEL INTO COLD WATER WHEN COOLING. ALLOW THE VESSEL
TO COOL DOWN NATURALLY.
A SUITABLE INSPECTION WARNING PERIOD OF 2000 FIRINGS, AND A STOP OPERATING LIMIT OF
2500 FIRINGS ARE PRE-PROGRAMMED FROM THE FACTORY
CAL3K-ST Installation Guide V 1.0
CAL3K Advanced Bomb Calorimeters
12/10/2023
Digital Data Systems (Pty) Ltd.
www.ddscalorimeters.com
Website:
www.ddscalorimeters.com
6