Operation; Filter, F-1001 F-10Ml - Rainin Pipetman F Series Manual

Single-volume microliter pipettes fixed volumes from 0.1 µl to 10 ml
Table of Contents

Advertisement

Operation

1. Attach a new disposable tip to the pipette shaft with
sufficient force to make a positive seal.
F-1MIC
F-1
F-200 use tips for volumes to 250 µL
F-201
F-1001
F-5001
2. Press the plunger to the
ed volume indicated on the plunger button.
3. Holding Pipetman upright, insert the tip into the sample
to the proper immersion depth:
1 to 2 mm for F-1MIC
2 to 4 mm for F-201
3 to 6 mm for F-1001
4. Allow the plunger button to return slowly to the up
position. Never let it snap up! Wait a moment so that
the full volume of sample is drawn into the tip.
5. Withdraw the tip from the sample. Carefully wipe any
liquid from the outside of the tip with a lint-free tissue,
taking care not to touch the tip orifice.
6. To dispense sample, touch the tip end against the side
wall of the receiving vessel and press the plunger slow-
ly to the
FIRST STOP
1 second* (F-1MIC
1-2 secs* (F-201
2-3 secs* (F-1001
Then press the plunger to the
stroke), expelling any residual liquid in the tip.
7. With the plunger fully depressed, withdraw Pipetman
from the vessel carefully, sliding the tip along the wall
of the vessel. Let the plunger return to the
8. Discard the tip by pressing the tip ejector button. Use a
fresh tip for each sample to prevent sample carryover.
Filter, F-1001
Models F-1001
shaft to help prevent liquid entering the shaft, particularly
important with large volumes. If the filter gets wet, replace
it. For F-1001 to 5000 insert the small diameter into the
shaft; for F-5001 to 10ML the large diameter (see below).
Part numbers are 6190-164 (pack of 100) and 6190-165
(pack of 1000).
2
F-10MIC use tips for volumes to 10 µL
F-1000 use tips for volumes to 1000 µL
F-5000 use tips for volumes to 5 mL
F-10ML use tips for volumes to 10 mL
. Wait:
F-10MIC, F-1
F-1000)
F-10ML) *Longer for viscous solutions.
F-10ML
F-10ML use a filter in the end of the
. This is the calibrat-
FIRST STOP
F-10MIC, F-1
F-1000
F-10ML
SECOND STOP
F-200
F-200)
(bottom of
position.
UP