6. Improper reassembly. Remove the tip ejector and
shaft. Check the position of the internal assemblies,
especially the seal, against the illustrations.
Consult the Rainin publication "Pipetman Care and Main-
tenance" (AB-14) for more information. This free publica-
tion is available by calling 800-543-4030 or it can be
downloaded in PDF format from Rainin's website:
http://www.rainin.com/pdf/ab14.pdf

Removing/Replacing Tip Ejector Arm

8
Tip Ejector Arm Removal
1. Press tip ejector button.
2. Pull arm from handle
using ejector arm release
collar.
Tip Ejector Arm Replacement
1. Press tip ejector button.
2. Hold Pipetman in one hand
and the tip ejector arm in the
other hand.
3. Align the plastic shaft at the
end of the tip ejector with the
rod inside the handle.
4. At the same time place the tip
ejector opening over the end
of the Pipetman shaft.
5. Push the plastic collar into
the handle until the tip ejector
is securely in place.