Chapter 4. Basic Programming
4.6.2 Alarm Type (cont.)
Users can configure the TMO2D alarms for either normal or failsafe mode. For normal alarm configuration, the alarm
contact remains de-energized until an alarm condition occurs. For failsafe mode, the alarm contact remains energized
until an alarm condition occurs, when it becomes de-energized.
ALARM FEATURES
Normal/Failsafe Mode?
Failsafe Alarm?
[NORMAL] failsafe
ALARM FEATURES
Normal/Failsafe Mode?
ALARM FEATURES
Done?
Select Relay to set:
[A]
B
C
D
done
MAIN MENU
Relays?

4.7 The Tests Menu

The fourth main menu, the Tests Menu, provides assistance in testing and troubleshooting the inputs and outputs of the
TMO2D.
The Tests Menu contains three submenus:
DVM Test
Output Test
Relays Test
(Refer to Figure 27 on page 80, for a flow diagram of the Tests Menu.)
34
Press [YES] to enter the normal/failsafe mode function.
Use the [NO] or arrow keys to indicate whether or not you wish the alarm to act as a
normal or failsafe alarm, and press [YES] to confirm the selection.
Press [NO] to exit the normal/failsafe mode function.
Press [YES] to leave the Alarms submenu.
Use the [NO] or arrow keys to select another relay, or select Done to leave the
Relays Menu. Press [YES] to confirm the selection.
Press [NO] to proceed to the next
menu — the Tests Menu.
TMO2D Digital Display User's Manual