Xmtc Transmitter - GE TMO2D User Manual

Panametrics display and control module
Table of Contents

Advertisement

Chapter 2. Installation
2.3.3b CE Mark Compliance (cont.)

2.3.4 XMTC Transmitter

Connecting the XMTC thermal conductivity transmitter to the TMO2D requires use of the X4(*) or Y4(*) cables. The
X4(*) cable can be used for distances up to 450 ft (130 m). For longer distances, each cable can be supplied as
18-AWG that can be located up to 1,050 ft (320 m). If you are using your own cable, refer to Table 7 below for cable
requirements.
For longer cable lengths, consult the factory.
12
1
2
3
4
Figure 7: TMO2-TC Wiring Connections
Table 7: Cable Requirements for XMTC
MAX. CABLE LENGTH
Feet
Meters
450
130
700
200
1,050
320
1,700
500
2,800
850
4,000
1,200
Connect Shield
Red
+24 V/1 AMP 1
Black
Return 2
White
Oxygen 3
None 4
CABLE SIZE
AWG
Sq. mm
22
0.35
20
0.6
18
1.0
16
1.2
14
2.0
12
3.0